HMI menu ID Parameter Description Default value
[unit]
Value range Access
level
2.3.2.5 Decimal Places 1) 2) 3) Define the decimal point for the
process value triggered in the
Local display.
2
● Auto
● No decimal
place
● 1
● 2
● 3
● 4
● 5
● 6
User
2.3.2.6 Direction Select totalizing direction. Forward count‐
ing
● Forward
counting and
backward
counting
● Forward
counting
● Backward
counting
● Hold
User
2.3.2.7 Fail-safe behaviour Select reaction of the totalizer
function in case of unreliable
process flow value.
Run
● Run
● Hold
● Memory
User
2.3.2.8 Reset Reset totalizer 2 to zero
● Cancel
● Ok
User
2.3.2.9 Preset Reset totalizer 2 to selected pre‐
set value
● Cancel
● Ok
User
2.3.2.10 Alarm and warning limits
2.3.2.10.11) Upper alarm limit Set upper alarm limit. Alarm is
triggered if totalizer 2 value is
above the limit.
[kg] User
2.3.2.10.21) Upper warning limit Set upper warning limit. Waning
is triggered if totalizer 2 value is
above the limit.
[kg] User
2.3.2.10.5 1) Hysteresis Enter the hysteresis to avoid un‐
wanted rapid switching.
0
[kg]
User
2.3.2.10.12) Upper alarm limit Set upper alarm limit. Alarm is
triggered if totalizer 2 value is
above the limit.
[m3] User
2.3.2.10.22) Upper warning limit Set upper warning limit. Waning
is triggered if totalizer 2 value is
above the limit.
[m3] User
2.3.2.10.32) Lower warning limit Set lower alarm limit. Warning is
triggered if totalizer 2 value is
above the limit.
[m3] User
Default settings
B.4 Totalizer
MASS 2100 & FC300 (FCT030)
Operating Instructions, 06/2017, A5E39789040-AA 287