SIREMOBIL Iso-C 3D SPR2-230.814.01 Page 6 of 14 Siemens AG
Rev. 05 08.04 CS PS 24 Medical Solutions
2 - 6 Actions up to serial number 1408
Upgrade cable set, item number 71 39 996 2
• Move the single wire (Fig. 14) on the right side from the top to the bottom by pulling it
through the horizonal carriage and the lifting column.
• Connect the single wire to the wire at the bundled cables (Fig. 13).
• Lay the cable set out on the floor (Fig. 13).
• Wrap the connections of the two wires using Scotch tape.
• Cover the tape with talcum powder, or use silicone or a similar type of lubricant.
• Pull the wire upward toward the top at the entrance to the horizontal carriage. You will
need one person pulling the wire upward and a second person feeding the wire from
below (Fig. 16).
• If necessary, push the existing bundle of cables in the recess of the horizontal carriage to
the left using a large screwdriver.
• In case of blockage, pull the cables back a little, turn them and continue pulling them
toward the top.
• Pull the cables up to the mark.
• Remove the two wires.
Fig. 13 Upgrade cable set Fig. 14 Single wire
Fig. 15 Wire on top Fig. 16 Wire below
right side