Functions
7.3 Logger Control
FUE1010 IP65 NEMA 4X
108 Operating Instructions, 12/2014, A5E03086491-AC
To send Logger contents to the display screen:
1. Press <Right Arrow> to access the [Display Logger] option list.
2. Scroll cursor to either [Line Wrap] or [No Line Wrap] by pressing <Up/Down Arrow>.
3. To view Logger contents press <ENTER>.
4. To return to [Logger Control] press <MENU>.
This menu cell allows you to send the Logger contents to an external device (usually a
computer or printer) via the flow meter’s RS-232 Serial I/O port. This command is effective
only after a successful install. In addition, you have to enable datalogging and select data
items in the [Logger Setup] menu.
The flowmeter interfaces with most serial printers or personal computers for Logger
printouts. You must use the proper cabling between the flowmeter and the external device.
In addition, you must configure the RS-232 Setup correctly. You should turn off the Logger
function before you transmit an extensive printout. This will avoid contaminating the printout
with new Logger data. Logger reports are sequential ASCII text files.
To send Logger contents to the RS-232 Serial Port:
1. Check the flowmeter-to-external device connections and your RS-232 Setup parameters
(see RS-232 Setup menu).
2. To access the [Output Logger] option list press <Right Arrow>.
3. Scroll the cursor to [Yes] by pressing <Up/Down Arrow>.
4. To transmit Logger contents to external device via the serial port press <ENTER>.
5. To stop printout press <Left Arrow>.