5.2 Transmitter Wiring
FUE1010 IP65 NEMA 4X
44 Operating Instructions, 12/2014, A5E03086491-AC
Notes on Clamp-on RTD Installation
The Clamp-On RTD sensors of the flow meter are extremely sensitive and precise. Their
contribution to the performance of your meter is every bit as important as that of the liquid
flow sensors. Please consult the installation drawings in Appendix B (Page 209) for details
on physical installation and wiring of the RTDs.
Black
TB1-T1 Channel 1 TS
Orange
TB2-T2 Channel 2 TR
Brown
To Sensor
Red
TB3-T1 Channel 1 TS
Blue
TB4-T2 Channel 2 TR
Figure 5-4 Dual Channel Temperature Sensor
Table 5- 1 TB3 and TB4 Wiring
Pin
TB3 Function
TB4 Function
Use
Description
Behavior
Load
Wiring
1
AUX. 1 IN
AUX. 3 IN
Iin1 Input
Analog current
input referenced
to meter
ground.
4 to 20mA 200Ω 1000 ft.
Max w/o
factory
approval
2
AUX. 1 COM
AUX. 3 COM
Iin1 Common
3
AUX. 2 IN
AUX. 4 IN
Iin2 Input
4
AUX. 2 COM
AUX. 4 COM
Iin2 Common