Connecting
5.4 Programming the Transmitter
FUE1010 IP65 NEMA 4X
54 Operating Instructions, 12/2014, A5E03086491-AC
Use this menu cell to enter the number of pipe diameters between the upstream configura-
tion and the sensor installation.
Use this menu cell to select the pipe configuration that most accurately represents the up-
stream pipe condition.
4. Press the <Left Arrow> and return to the main menu.
Table 5- 3 Pipe Configuration Option List Definitions
Options
Definitions
Fully Developed
Fully developed flow, as would be expected for very long
straight pipe runs or installation downstream of a flow condi-
tion.
1 Elbow
Single 90 degree Elbow upstream of sensor installation.
Dble Elbow+
Double out-of-plane Elbows upstream of sensor installation.
Dble Elbow-
Double in-plane Elbows upstream of sensor installation.
Valve
Not available at this time.
Expander
Pipe expansion upstream of sensor installation.
Reducer
Pipe reduction upstream of sensor installation.
Norm Entry
Not available at this time.
Header Inlet
Header or pipe manifold upstream of sensor installation.
Intrusions
Not available at this time.