Functions
7.12 Reflexor
FUE1010 IP65 NEMA 4X
138 Operating Instructions, 12/2014, A5E03086491-AC
3. Adjust the Noise Level cursor, if required, by pressing the <+> key to increase or <-> key
to decrease.
Note
A delay in response to all keys will occur. This is due to the longer periodic sampling of
the keyboard by the processor while performing FFT’s.
4. Press the <Down Arrow> from Noise Level to advance to the High Limit cursor control.
The <+> and <-> key are active in this position. Refer to the "Available Adjustments To
Spectra Graph" table for details.
5. Press the <Down Arrow> from High Limit to advance to the Low Limit cursor control. The
<+> and <-> keys are active in this position. Refer to the "Available Adjustments To
Spectra Graph" table for details.
6. Press the <Down Arrow> from Low Limit to advance to the Carrier FX control. The <+>
and <-> keys are active in this position.
7. Press the <Down Arrow> from Carrier FX to advance to the Diagnostic Data Display. The
<+> and <-> keys are active in this position.
8. To exit the Spectra Graph, press the <MENU> key.
9. To display the Digital Flow Display, press the <MENU> key from any Installation Menu
location.
Available Spectra Graph Adjustments
Noise Level Set Noise Filter Set to approximately one quar-
ter the Peak FFT value for typi-
High Limit Low Pass Filter Setting Set to two times the maximum
Low Limit High Pass Filter Set to one half minimum flow if
noise is present. In most cases,
Carrier Fx Transmit Frequency Code Adjust only under guidance of
The [Diagnostic Data] menu and the [Application Info] sub menu screen provide one location
where all diagnostic data can be viewed.
Note
Adjustments can not be made on the [Application Info] menu screen. Adjustments can only
be made on the Spectra Graph.