EasyManua.ls Logo

Siemens SITRANS FUE1010 IP65 NEMA 4X - Page 59

Siemens SITRANS FUE1010 IP65 NEMA 4X
296 pages
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
Connecting
5.5 Sensor Installation
FUE1010 IP65 NEMA 4X
Operating Instructions, 12/2014, A5E03086491-AC
57
Upstream Sensor
Downstream Sensor
Pipe
Sonic path
Flow Direction
Figure 5-8 Direct Mount (Pipe shown from above in 12 o'clock position)
Mounting Supplies
The following items will be needed to mount the sensors (most are supplied):
Flat blade screwdriver
Mounting Frames or Mounting tracks
Tape, chalk and a ruler or measuring tape
Mounting Straps
Spacer Bar
Mounting Guide (for Direct Mount)
Ultrasonic coupling compound and/or coupling pads
Sensors (matched set)
Mounting Strap Kits
The available Mounting Strap kits are listed below. Each kit comes with up to two band sizes
to cover its designated pipe diameter range and a spacing guide for Direct Mount.
Mounting Strap Kits
Pipe Diameter
SAE Band Sizes (Qty.)
7ME396000SM00
50.8mm (2-inch) to 177.8mm (7-inch)
#88 (2) #128 (2)
7ME396000SM10
50.8mm (2-inch) to 330.2mm (13-inch)
#88 (2) #152 (2)
7ME396000SM20 330.2mm (13-inch) to 609.6mm (24-
inch)
#188 (2) #280 (2)
7ME396000SM30 609.6mm (24-inch) to 1219.2mm (48-
inch)
#152 (4) #312 (4)

Table of Contents

Related product manuals