10.3 Other technical data
a) type of insulation ................................................double, according to EN 61010-1 and IEC 61557
b) measurement category ...................................... IV 300 V (III 500 V) according to EN 61010-2-030
c) housing degree of protection for the housing acc. to EN 60529 ....... IP51 (with closed sealing plug)
d) meter power supply ........................................................................... Li-Ion 11,1 V 3,4 Ah 37,7 Wh
e) parameters of AC adapter for the battery charge .................................................... 12 V DC / 2.5 A
............................................................................................. 100 V…240 V, 50 Hz…60 Hz (mains)
f) dimensions ..........................................................................................288 mm x 223 mm x 75 mm
g) weight of the meter with batteries ............................................................................. approx. 2.5 kg
h) storage temperature .................................................................................................. -20°C...+60C
i) operating temperature .................................................................................................. 0°C...+45C
j) temperature range suitable for initiating battery charging.......................................... +10C...+40C
k) temperatures at which the charging process is interrupted .............................. <+5 C and ≥ +50C
l) humidity ......................................................................................................................... 20%...90%
m) reference temperature ................................................................................................. +23°C ± 2C
n) reference humidity ......................................................................................................... 40%...60%
o) altitude (above sea level): ................................................................................................. <2000 m
p) time until Auto-OFF ............................................................................................ 2 min, 5 min or off
q) number of measurements Z or RCD (for batteries) ...................... >3000 (6 measurements/minute)
r) number of measurements R
ISO
or R (for charged batteries) .................................................. >1000
s) recording time (for batteries) .................................................................................................... 16 h
t) display .............................................................................................. colour LCD TFT, touchscreen
.............................................................................................................................. 800 x 480 pixels
...................................................................................................................................... diagonal 7"
u) memory of measurement results ...................................................................................... unlimited
v) recorder memory .............................................................................................................. unlimited
w) data transmission .................................................................................................................... USB
x) quality standard design, construction and manufacturing are ISO 9001, ISO 14001, PN-N-18001
compliant
y) the device meets the requirements of IEC 61557 standard
z) the product meets EMC requirements (immunity for industrial environment) according to the fol-
lowing standards..............................................................................................................................
........................................................................................................ EN 61326-1 and EN 61326-2-2