MPI-540 USER MANUAL
69
Read out the result.
Signal lights for reaching the limit
(section 3.6.1 step ):
the result is within the set
limit
the result is outside the set
limit
assessment not possible
After selecting the bar on the
right side, a menu will slide out
with additional measurement re-
sults.
R
H
resistance of current elec-
trode
R
S
resistance of voltage elec-
trode
δ additional uncertainty caused
by resistance of the electrodes
Selecting the bar hides the
menu.
Save the measurement to the meter memory using the icon. A detailed description of
memory management is contained in section 5.1.3.
The last measurement can be recalled using the icon.
WARNING
Measurement of resistivity may be carried out if voltage of interferences does
not exceed 24 V. Voltage of interferences is measured up to the level of 100 V.
Over 50 V it is signalled as hazardous. The meter must not be connected to
voltages exceeding 100 V.