R6,Calibration
80
11 Calibration
11.1 Overview
Description
Stonex
®
instrumentsaremanufactured,assembledandadjustedtoahighquality.Quick
temperaturechanges,shockorstresscancausedeviationsanddecreasetheinstrumentaccu‐
racy.Itisthereforerecommendedtocalibratetheinstrumentfromtimetotime.Thiscanbe
doneinthefieldbyrunningthroughspecificmeasurementprocedures.Theproceduresare
guidedandhavetobefollowedcarefullyandpreciselyasdescribedinthefollowingchapters.
Someotherinstrumenterrorsandmechanicalpartscanbeadjustedmechanically.
Electroniccalibration
Thefollowinginstrumenterrorscanbecheckedandcalibratedelectronically:
Horizontalcollimationerror,alsocalledline‐of‐sighterror.
Verticalindexerror,andsimultaneouslytheelectroniclevel.
)
Fordeterminingtheseerrors,itisnecessarytomeasureinbothfaces,buttheprocedurecanbe
startedinanyface.
Mechanicalcalibration
Thefollowinginstrumentpartscanbecalibratedmechanically:
Circularlevelontheinstrumentandtribrach.
•Laserplummet.
Screwsonthetripod.
)
Duringthemanufacturingprocess,theinstrumenterrorsarecarefullydeterminedandsetto
zero.Asmentioned,theseerrorscanchangeanditishighlyrecommendedtoredeterminethem
inthefollowingsituations:
Beforetheinstrumentisusedforthefirsttime.
Beforeeveryhighprecisionsurvey.
Afterroughorlongperiodsoftransport.
Afterlongperiodsofworkorstorage.
Ifthetemperaturedifferencebetweencurrentenvironmentandthetemperatureatthe
lastcalibrationismorethan10°C(18°F).
11.2 Preparation
)