83
PPoowweerr
Control System will not
power up
Check to see if power cable is firmly connected
Ensure front power switch is ON
Touch panel will not power
up
Check to see if power cable is firmly connected
Unplug the power cable and plug it in again
Verify the green LED light located on the back
underside of the touch panel is lit, indicating the
monitor is ON
AAuuddiioo
SSyymmppttoomm
AAccttiioonn
No audio in room
Check that Speaker Controls volume is turned up and is
not muted
Audio distorted Decrease the input level
Turn down the output level of speaker level
No audio from surgeon’s
microphone
Make sure microphone is turned ON
Check batteries and replace if necessary (
see 27.5.1
Replacing the Wireless Microphone Batteries, page
99
)
Audio not leaving room Check the wireless microphone batteries if the mic is
not heard in other locations. The LED on the top of
the wireless mic belt pack should light up
momentarily when powered ON and then go out. If it
stays on, the battery needs to be replaced (
see
27.5.1 Replacing the Wireless Microphone Batteries,
page 99
)
Check the wireless receiver. The front panel on the
wireless receiver should display RF and AF level bars
when talking into the microphone
Verify that mic frequency matches the receiver
frequency
VViiddeeoo
SSyymmppttoomm
AAccttiioonn
No image in preview when
source is selected
Ensure S-video cable is connected firmly to the back
of the Control Section and Output 2 on the router
Ensure that the video source is turned ON
For AIO systems, make sure that the DVI cable is fully
connected to the AIO touch panel and to Output 2 on
the router.