Appendix
MA6 / MA8 - 3 Appendix - Rev.05 07-06
91
3
RESET-STEPS
1. TSA Y-axis
2. TSA X-axis
3. Stage Z-axis
4. BSA X-Y-axis
(Y-left, Y-right, wait for S6.2 + S6.4, X-left, X-right, wait for S6.1 + S6.3, move to default position)
5. Motorized alignment stage if applicable / (Separation Knife if applicable)
6. Main Menue
SERVICE PROGRAM LEVEL
WARNING! ONLY FOR SKILLED STAFF
Press ALIGN/CONT/EXP-key simultaneous to switching on machine!
1. VALVE TEST
2. INPUT TEST
3. LED TEST
4. SET GRAB Z POSITION
5. TSA STAGE
6. BSA STAGE
7. ALIGNMENT STAGE
8. HEIDENHAIN
9. I/O Test (with special equipment only)
Reset the machine: press FAST key first and then UNLOAD key second within 200 ms.