Appendix C:  Performance Verification
Table A-8:  Deflection Accuracy Limits
VOLTS/DIV
Switch
Setting
Standard
Amplitude
Signal
Accuracy
Limits
(Divisions)
2 mV 10 mV
4.90  to  5.10
5 mV 20 mV
3.92  to  4.08
10 mV 50 mV
4,90  to  5.10
20 mV
0.1  V
4,90  to  5.10
50 mV
0.2 V
3.92  to  4.08
0.1  V
0.5 V
4.90  to  5.10
0.2 V
1 V
4.90  to  5.10
0.5 V
2 V
3.92  to  4.08
1  V
5 V
4.90  to  5.10
2 V
10 V
4.90  to  5.10
5 V
20 V
3.92  to  4.08
c.  Move the cabie from the CH 1  OR X input connector to the CH 2 OR 
Y input connector. Set the Vertical MODE switch to CH 2.
d.  Repeat part b using the CH 2 controls.
□   Step 2:  Check Store Deflection Accuracy
a.  Set:
CH 2 VOLTS/DiV  2 mV
STORE/NON-STORE  STORE (button  in)
Acquisition MODE  AVERAGE
b.  Set the generator to produce a five division standard amplitude 
signal.
c.  Use the CURSORS control and SELECT C1/C2 switch (push in the 
CURSORS controls knob) to set one cursor at the bottom of the 
square wave and the other cursor at the top of the square wave.
d.  CHECK — Deflection accuracy is within the limits given in Table A-9 
for each CH 2 VOLTS/DIV switch setting and corresponding stan
dard-amplitude signal.
e.  Move the cable from the CH 2 OR Y input connector to the CH  1  OR 
X input connector. Set the Vertical MODE switch to CH 1.
f.  Repeat parts b and c using the CH 1  controls.
A-30
Appendices