Saving Waveforms and Setups
P I  Step 3:  Press the Select Setup menu button to select either Setupl  or 
Setup2 memory location for the setup information,
H   Step 4:  Press Save Setup to save to current acquisition and menu 
settings to the selected memory location.
Recalling a Setup from Memory
Recall a saved setup either by returning to the ADV FUNCT/Save Setup 
Menu or by specifying the Power Up State as the particular setup you want 
to recall. (Refer also to the following section, Recalling a Power Up State.)
P I  Step 1:  Press the Setup ADV FUNCT button. The Advanced Functions 
menu appears.
F I   Step 2:  Press the button underneath Save Setup Menu.
F I  Step 3:  Press the Select Setup menu button to select either Setupl  or 
Setup2 memory location,
F I  Step 4:  Press Recall Setup to recall the acquisition and menu settings 
previously saved to that memory location.
Recalling a Power Up State
The osciiioscope can be configured to return to one of four setups when the 
instrument is turned on.
F I  Step 1:  Press the Setup ADV FUNCT button. The Advanced Functions 
menu appears.
n   Step 2;  Press the button underneath Save Setup Menu.
f l   Step 3:  Press the button underneath Power Up State to select one of 
the following alternatives:
Default — The instrument will use the factory default settings when instru
ment powers up,
Pwr Dwn — The instrument will automatically save the software-controlled 
settings when the instrument is turned off. The Instrument will return to these 
same settings when the instrument is turned back on,
Setupl — The instrument will use the setup saved under Setupl  at power 
up.
Setup2 — The instrument will use the setup saved under Setup2 at power 
up.
f l   Step 4:  The selection is now saved. Press the ADV FUNCT button to 
return to normal operation.
2221A User Manual
3-43