iv i ensuring otgnais
it^
AVI
= 81. 0 *
TR
:g i =
B2*  .  AT = 17
2 .0 r
15
pT  " 
Si/
“  -  r  -
/ ir ~
\
XTw
|
s
"v
-  _
-  -
-  ■
E
> 0.
5U
su
ERfiGE  5 ^ n s
Figure 3-38:  Measuring Fail Time
CD Step 1:  Apply a signal to the oscilloscope.
n   Step 2:  Set the trigger SLOPE out (positive) for a rise time measure
ment and in (negative) for a fall time measurement.
F I  Step 3:  Adjust the vertical dimension for exactly 5 divisions (you may 
have to use the CAL variable.)
n   Step 4:  Vertically position the signal so that the bottom of the signal on 
the 0% graticule line and the top of the signal is on the 100% line.
□   Step 5:  Magnify the rising edge of the signal horizontally so that the rise 
time is spread over 4 or 5 divisions. (Refer to Displaying Magnified 
Sweeps, page 3-19.)
F I  Step 6:  Measure the rise time horizontally from the 10(%) graticule line 
to the 90(%) graticule line,
Measuring Trigger Level
The trigger level is the point on the rising or falling edge of a signal where 
the oscilloscope triggers a sweep. This voltage level is indicated by the TRIG 
readout at the top of the CRT. TRIG1  represents the level of the trigger signal 
coming from Channell  and TRIG2 represents Channel 2.
Measuring Phase Difference
With the two vertical channels on the oscilloscope you can measure phase 
differences on signals that range in frequency anywhere within the limits of 
the vertical system. Use the following procedure;
f~l  Step 1:  Set both input coupling switches to the same position, depend
ing on the type of input coupling desired.
3-38
tn Detail