Appendix C: Performance Verification
nf
j.  Press in the SAVE/CONT button to select CONT.
k.  Position the trace 0.5 division above the bottom horizontal graticule 
line using the CH 2 POSITION control.
l.  Press in the SAVE/CONT button to select SAVE.
m.  CHECK — Trace shift of 0.5 division or less.
n.  Press in the SAVE/CONT button to select CONT.
o.  Set the Vertical MODE switch to CH 1.
p.  Repeat steps g through m for CH  1 trace.
□   Step 6:  Check Bandwidth
a.  Set:
VOLTS/DiV (both) 
AC-GND-DC (both) 
SEC/DIV 
BW LIMIT
STORE/NON-STORE
2 mV 
DC
0.2 ms
Off (button out) 
NON-STORE (button 
out)
b.  Connect the leveled sine wave generator output via a 50 Q precision 
cable and a 50 Q termination to the CH 1  OR X input connector.
c.  Set the generator to produce a 50 kHz, six division display.
d.  CHECK — Display amplitude is 4.2 divisions or greater as the gener
ator output frequency is increased up to the value shown in Table 
A-10 for the corresponding VOLTS/DIV switch setting.
Table A-10:  Settings for Bandwidth Checks
VOLTS/DIV
Generator
Switch Setting Output Frequency
2 mV
80 MHz
5 mV to 0.5 V
100 MHz
e.  Repeat parts c and d for all indicated CH 1  VOLTS/DIV switch 
settings, up to the output-voltage upper limit of the sine wave gener
ator being used.
f.  Move the cable from the CH 1 OR X input connector to the CH 2 OR 
Y input connector.
g.  Set the Vertical MODE switch to CH 2.
h.  Repeat parts c and d for all indicated CH 2 VOLTS/DIV switch 
settings, up to the output-voltage upper limit of the sine wave gener
ator being used.
I
2221A User Manual
A-33