Appendix C: Performance Verification
H Step 7:  Check Repetitive Store Mode and Bandwidth
a.  Set:
CH 2 VOLTS/DIV  10 mV
SEC/DIV  0.2 ms
b.  Set the generator to produce a 50 kHz, six division display.
c.  Set:
SEC/DIV  0.05 ps
X10 Magnifier  On (knob out)
d.  Set the generator to produce a 100 MHz display.
e.  Set:
STORE/NON-STORE  STORE (button in)
SAVE/CONT  CONT
NOTE
Allow the points to accumulate for a few seconds before saving the 
display.
f.  Press in the SAVE/CONT button to select SAVE.
g.  CHECK — The 100 MHz display is saved.
h.  CHECK — Display amplitude is 4.2 divisions or greater.
i.  Press in the SAVE/CONT button to select CONT.
j.  Set the Vertical MODE switch to BOTH and ALT.
k.  Repeat parts 1 through h.
["I  Step 8:  Check Single Sweep Sample Acquisition
a.  Set:
Vertical MODE 
SEC/DIV 
X I0 Magnifier 
Trigger Mode 
SOURCE 
SAVE/CONT
b.  Set the generator to produce
c.  Press in the Trigger Mode SG
CH 2
5 ps
Off (knob in) 
NORM 
CH 2 
CONT
50 kHz, six division display. 
SWP button.
d.  Set the generator output to 2 MHz.
e.  Press in the Trigger Mode SGL SWP button.
f.  CHECK — the minimum peak-to-peak envelope amplitude is greater 
than 5.6 divisions.
A-34
Appendices