EasyManuals Logo
Home>Tektronix>Test Equipment>2221A

Tektronix 2221A User Manual

Tektronix 2221A
170 pages
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Page #95 background imageLoading...
Page #95 background image
Measuring Signals
C l Step 2: Using either probes or coaxial cables with equal time delays,
connect a known reference signal to the Channel 1 input and the un
known signal to the Channel 2 input.
f l Step 3: Switch the Vertical Mode to BOTH and then select either ALT or
CHOP
H Step 4: Set both VOLTS/DIV switches and both variable controls so the
displays are equal in amplitude.
f l Step 5: Set the SOURCE to CH 1 so the oscilloscope uses only the
reference signal for triggering. Adjust the TRIGGER LEVEL control for a
stable display.
f l Step 6: Set the SEC/DIV switch to a sweep speed that displays about
one full cycle of the reference waveform.
f~l Step 7: Position the displays and adjust the SEC/DIV variable control so
that one cycie of the reference signal occupies exactly 8 horizontal
divisions. Each horizontal division of the graticule now represents 45° of
the cycle (360° 8 divisions).
F l Step 8: Measure the horizontal difference in divisions between the
signals and multiply it by 45° (Figure 3-39).
Channel 1 Reference
Channel 2
Figure 3-39: Measuring Phase Difference
For higher resolution measurements of the phase difference the display may
be magnified horizontally by pulling the XI0 magnification knob out
(Figure 3-40).
2221A User Manual
3-39

Table of Contents

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Tektronix 2221A and is the answer not in the manual?

Tektronix 2221A Specifications

General IconGeneral
BrandTektronix
Model2221A
CategoryTest Equipment
LanguageEnglish

Related product manuals