Checking the Probe Compensation
NOTE
Always compensate a probe for the particular channel that you use 
it with.
Use the following procedure to check the probe compensation:
Step 1:  Set the instrument controls as described in Setting up the 
Display.
n   §tep 2:  Connect the probe to the channel you intend to use it for.
{If the probe is properly “coded” the volts-per-division readout for the chan
nel will change to match the attenuation factor of the probe. The 10X probes 
supplied with the oscilloscope already have the proper coding.)
□   SteP 3:  Clip the probe tip to the PRB ADd connector.
f~|  Step 4:  Use the VOLTS/DIV control to display about 5 divisions of the 
waveform,
□   Step 5:  Use the VERTICAL POSITION and HORIZONTAL POSITION
controls to center the display.
F I  Step 6:  Check the waveform against Figure 1-3 to see if the probe is 
correctly compensated.
r
Over compensated
r
r
v .
Under compensated
Correctly compensated
Figure 1-3:  Checking Probe Compensation
f~l  Step 7:  Adjust the compensation if necessary.
7-6
Before You Begin