Commands
5-18
TDS5000B Series Service Manual
quite common, however, to use uncalibrated probes (particularly when the
instrument inputs are connected into a test system with coax cables).
This command is equivalent to selecting Instrument Calibration from the
Utilities menu, and then viewing the contents of the Status field.
CALibrate:RESults?
CALIBRATE:RESULTS?
This query returns the status of all the calibration subsystems. It might return:
:CALIBRATE:RESULTS:SPC PASS;FACTORY PASS;FACTORY:VOLTAGE
PASS;FREQUENCY PASS;TRIGGER PASS
CALibrate:RESults:SPC?
This query-only command returns the results of the last SPC operation: either
PASS or FAIL.
*CAL?
CALibrate:RESults:SPC?
CALIBRATE:RESULTS:SPC?
Returns the results of the last SP C operation.
CALibrate:TEMPerature?
The query returns the air temperature inside the case in degrees Celsius.
CALibrate:TEMPerature?
CALIBRATE:TEMPerature?
This query might return
:CALIBRATE:TEMPERATURE 20
indicating that the air temperature inside the case is 20 degrees Celsius.
Syntax
Example
Related C ommands
Syntax
Example
Syntax
Example