Removal and Installation Procedures
TDS5000B Series Service Manual
6--11
c. Slide the right side panel towards the rear of the oscilloscope allowing
the tabs to clear the cover openings, then pull out to remove the panel
from the oscilloscope.
4. Remove the VGA panel: Use Figure 6--2, page 6--13 as a guide.
a. Remove the VGA panel by gently pulling the trim up and out from the
oscilloscope.
5. Remove the pouch: Use Figure 6--2, page 6--13 as a guide
a. Unsnap the pouch from the top cover.
b. Remove the two T-15 Torxdrive screws that secure the power cord
retainers and pouch bracket at the top rear of the oscilloscope.
6. Remove the top cover: Use Figure 6--2, page 6--13 as a guide.
a. Remove the accessory tray door or printer door.
b. Remove the accessory tray or internal printer (two T-15 Torxdrive
screws).
c. If the internal printer is installed, disconnect the cable from the printer.
d. Slide the top cover towards the rear of the oscilloscope allowing the tabs
to clear the cover openings, then pull out to remove the cover from the
oscilloscope.
7. Remove the front panel trim:
a. To prevent the power button from falling out of the front panel trim,
place a piece of tape over the button.
b. Remove the three T-15 Torxdrive screws that secure the front panel trim
to the oscilloscope.
c. Grasp the trim ring by its top edge and pull toward you to detach the
three plastic snaps. (Alternatively, you can use a flat-bladed solder aid or
other small prying tool to help you detach the snaps.)
d. Grasp the trim ring by its side edges and slide it back and forth to release
the snap in the middle of the trim ring. The snap is at the bottom “T” of
the front panel above the channel 1 and 2 BNCs.
e. Pull off the trim ring. As you do so, take care not to lose the 3 pem
inserts in the bottom of the trim ring.