92
The indirect reference is located below the level of the instrument base
1. In ST1 or ST2 mode, the user has to press first, in order to manually enter the
preset value.
2. The following step is the definition of the measurement reference. In our example,
this reference value is only a fixed point used to define the indirect reference.
Once steps 1 and 2 are finished, all measurements will be calculated according to the
indirect reference A.
The option can only be selected when the software requires to define a
measurement reference. Any modification of the value of the indirect
reference will implicate a redetermination of the measurement reference.
For further details, see this chapter.
14.8 A&B references
management
The MICRO-HITE and MICRO-HITE+M offer the possibility to work with two measurement
references called A and B at any moment. You will automatically be asked to define a
reference in the following situations:
Mode
Description
ST1
When you enter into this mode for the first time
ST2
Following a calibration of the probe
At the same time, it is also possible to force the definition or re-definition of a measurement
reference by:
pressing the key , which will restart the process of determining the active reference
pressing the key or on the control panel during 3 seconds. As a result, the last
measurement block in the measurement memory (calculated or measured) is taken into
account as active reference value.
A reference can be activated by pressing the key or on the control panel. This is
only possible if the reference value has been defined previously. Otherwise, an error beep
will warn you that this reference cannot be used for the moment as it has not been defined
yet.
14.9 Delete last
measurement
block
It is possible to delete the last measurement block in the memory at any moment via the
key . This deletion is independent from the selected block(s) in the measurements list.
14.10 Edit a
measurement
block
The name of a selected measurement block in the history can be edited at any time

Table of Contents

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the TESA MICRO-HITE+M and is the answer not in the manual?

TESA MICRO-HITE+M Specifications

General IconGeneral
BrandTESA
ModelMICRO-HITE+M
CategoryMeasuring Instruments
LanguageEnglish

Related product manuals