Measure-
ment
para-
meter
Concen-
tration
[ppm]
Recom-
mended
measure-
ment
period
[min]
Recom-
mended
rinse time
[min]
Recom-
mended
calibration
cycle in
months
Filter service life
NO
low
10
20
50
100
200
300
90
60
30
20
10
10
5
5
5
10
10
20
3
3
3
3
3
3
approx. 40.000ppmh
NO
2
10
20
50
100
200
500
90
60
30
20
10
10
5
5
5
10
10
20
3
3
3
3
3
1
-
SO
2
50
100
200
500
1000
2000
5000
90
60
30
15
10
10
5
5
5
10
10
10
20
40
3
3
3
3
3
1
1
approx.
200.000ppmh
H
2
S
10
20
50
100
200
300
40
30
20
10
5
5
5
5
10
10
10
20
2
2
2
2
2
2
-
CxHy
Pellistor
no rinsing cycles required, as long
as the flue gas contains a sufficient
amount of O
2
(O
2
shut-down...)
2
approx. 70.000ppmh
CO
2
-(IR) no rinse cycles required 1 -
If the testo 350 is not used for measurements over an
extended period, but rather, for example, for random
measurements during start-up, servicing and adjustment of
industrial combustion systems, process systems, power
plants, gas turbines or stationary industrial motors, an
annual check of the testo 350 by testo service is
recommended.