38
Cable length: max. 164 ft. between control unit and analyzer
box, max. 2,624 ft. between all analyzer boxes in the data bus
system.
The bus system must have a defined electrical termination, see
below.
Electrical termination of the bus system
The data bus system is linear in structure. The control unit or the
testo data bus controller with USB connection represents the
beginning of the line.
The end is represented by the last components connected in the
system (analyzer box or analog output box). This component must
have a defined electrical termination.
An analog output box is the furthest subscriber.
> Plug the data bus termination plug into the data bus socket on
the analog output box.
An analyzer box is the furthest subscriber.
> Set the data bus terminating slide switch on the analyzer box
(see Connections / interfaces page 27, point 1) to switch
position right (
).
5.2.4.3. Connection via Bluetooth (option)