TESTO 350 - Accuracy and Response Time

TESTO 350
108 pages
Print Icon
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
14
Measurement
parameter
Measurement
range
Resolution
Abs. Press.,
optionally when IR
sensor is installed
-240 to 461 "H
2
O 0.4 "H
2
O
Flow velocity 0 to 131 ft/sec 0.1 ft/sec to 131 ft/sec
Type K (NiCr-Ni) -328° to 2498 °F 32.18 °F
Type S (Pt10Rh-Pt) 0° to 3200 °F 33.8 °F
3.2.5. Accuracy and response time
Analysis box
Measurement
parameter
Accuracy
Response
time
O
2
±0.2Vol.% < 20s (t95)
CO, H
2
-comp. ±10ppm (0…199ppm)
±5% of reading (200…2000ppm)
±10% of reading (rest of range)
< 40s (t90)
CO
low
, H
2
-comp. ±2ppm (0…39.9ppm CO)
±5% of reading (rest of range)
< 40s (t90)
NO ±5ppm (0…99ppm)
±5% of reading (100…1999ppm)
±10% of reading (rest of range)
< 30s (t90)
NO
low
±2ppm (0…39.9ppm)
±5% of reading (rest of range)
< 30s (t90)
NO
2
±5ppm (0…99.9ppm)
±5% of reading (rest of range)
< 40s (t90)
SO
2
±5ppm (0…99ppm)
±5% of reading (100…1999ppm)
±10% of reading (rest of range)
< 30s (t90)
H
2
S ±2ppm (0…39.9ppm)
±5% of reading (rest of range)
< 35s (t90)