86
2. Unlock the condensate trap/condensate container and pull it
vertically off the analyzer box.
3. Open the drain plug (1) and let the condensate run out into a
sink.
4. Wipe off any drops still on the condensate outlet with a cloth
and close the condensate outlet.
5. Plug the condensate trap/condensate container on the analyzer
box.
7.10. Checking / replacing the dirt filter
Checking the dirt filter:
> Check the dirt filter of the analyzer box periodically for
contamination. Check visually by looking through the window of
the filter chambers. In case of visible contamination, change the
dirt filter.
Replacing the dirt filter:
The filter chamber may contain condensate.