6.1.4.6. Calibration data
With this function the current calibration data and the sensor status
of the individual sensors can be displayed.
The condition of the sensor is checked with each sensor
calibration / adjustment. The graphic representation shows the last
25 calibrations.
Activating the function:
> [
] Sensor settings [OK] Calibration data [OK].
Options
> [Options] [Print]: The current calibration data of all sensors
are printed out.
> [Options] [Graphic]: The status of the selected sensor is
graphically displayed.
Threshold Explanation
100% Full capacity
70%
Reduced sensor sensitivity.
Recommendation: Acquire a replacement
sensor
50% Replace sensor
6.1.4.7. Negative value
The display for negative values can be activated/deactivated.
Activating the function:
> [ ] Sensor settings [OK] Negative values
Switching negative values on/off
1. [Change]
2. Select setting: [], []
3. Confirm the entry: [OK]