5.5.1. Error diagnosis
> Error diagnosis [OK].
- Unrectified errors, warnings and notes are displayed.
> View next / previous error: [], [].
5.5.2. Gas path check
(only available via Analyzer Box tab)
Check the analyzer regularly for leaks, to ensure accurate
measurements.
The leak test requires a plastic cap 0193 0039, comes with the flue
gas probe).
1. Gas path check [OK]
2. Place the plastic cap on the tip of the flue gas probe so that the
openings are completely covered.
- The pump flow is displayed.
- Volumetric flow rate less than or equal to 0.04l/min: The gas
paths are leak tight (traffic light in display lights green).
- Volumetric flow rate higher than 0.04l/min: The gas paths are
leaking (traffic light in display lights red). Probe and analyzer
box must be checked for leaks.
5.5.3. Sensor diagnosis
(only available via Analyzer Box tab)
1. Sensor diagnosis [OK].
2. Select sensor. [], [].
- The status of the sensor is indicated by a lamp.
A sensor is able to recover. It is therefore possible that the
sensor status indication changes from yellow to green or
from red to yellow.
5.5.4. Instrument information
> Device information [OK].
- Information is displayed.