6.1.2.7. Printer
This function is available in both the analyzer box and the control
unit. This function is available for both the control unit and the
analyzer box.
The headers (lines 1-3) and the footer for the printout can be set.
The printer that is used can be activated.
Activating the function:
> [
] Device Settings [OK] Printer [OK]
Activating the printer:
The printer 0554 0543 can only be selected after the
Bluetooth-interface has been activated, see Bluetooth,
page 55.
1. Select Printer [OK].
2. Select the printer: [], [] [OK].
- The printer is activated and the menu Printer is opened.
Setting the print text:
1. Print text [OK].
2. Select function: [], [] [Edit].
3. Enter values [Next].
4. Save the entry: [Finished].
6.1.2.8. Bluetooth
This menu is only available if the instrument is equipped with
Bluetooth option. The Bluetooth module can be switched on / off.
This function is available in both the analyzer box and the control
unit. Settings only apply for the device activated at the time.
To set up a connection between Control unit and analyzer box, see
Connection via Bluetooth (option), page 38.
To set up a connection between control unit and Notebook/PC:
Follow the operating instructions for the software and Notebook/PC
used.
Calling up the function:
> [
] Device settings [OK] Bluetooth [OK].
w ww . . co m
information@itm.com1.800.561.8187

Table of Contents

Other manuals for TESTO 350

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the TESTO 350 and is the answer not in the manual?

TESTO 350 Specifications

General IconGeneral
Measurement ParametersO2, CO, NO, NO2, SO2, H2S, CO2, CxHy, H2, Temperature, Draft, Pressure
O2 measurement range0 to 25 vol. %
CO measurement range0 to 10, 000 ppm
SO2 measurement range0 to 5, 000 ppm
Resolution H21 ppm
Accuracy O2±0.2 vol. %
DisplayGraphic display
CommunicationUSB, Bluetooth
Power SupplyRechargeable battery
Operating Temperature-5 to 45 °C
Accuracy CO±5 ppm or ±5% of measured value
Accuracy NO±5 ppm or ±5% of measured value
Accuracy NO2±5 ppm or ±5% of measured value
Accuracy SO2±5 ppm or ±5% of measured value
Accuracy H2S±5 ppm or ±5% of measured value
Accuracy H2±5 ppm or ±5% of measured value
Draft measurement range-99.9 to +99.9 hPa