Question Possible causes/solution
Pump volumetric flow rate
Too low gas flow (filter clogged) or too
high gas flow (overpressure).
> Check gas path / filter.
Gas cooling system Gas cooler not working (faulty).
> Please contact your local dealer or
the Testo Customer Service.
Sensor temperature too
high
O
2
sensor temperature beyond the
specification.
Gas cooler fault
Condensate in gas cooler was not
pumped off.
> Check peristaltic pump
Inlet and outlet hoses mixed up by
mistake.
> Push the inlet and outlet hoses
correctly over the sockets on the
pump head.
Low pump power
Inlet and outlet hoses mixed up by
mistake.
> Push the inlet and outlet hoses
correctly over the sockets on the
pump head.
Slow to establish a
connection or slow data
transfer between
PC/notebook and testo 350
via Bluetooth
Control unit is plugged into measuring
box or is connected via data bus cable.
> To achieve a high data speed, we
recommend setting up a Bluetooth
connection directly via the measuring
box.