Operation
Service Menu
Thermo Electron Corporation Model 48i Instruction Manual 3-115
Note Current channels are only displayed if I/O expansion board option is
installed.
• In the Main Menu, choose Service > Analog Out Calibration > Voltage
Channel 1-6 or Current Channel 1-6.
• Use the and to move the cursor up and down.
• Press to select a choice.
• Press to return to the Service menu or to return to the
Run screen.
WARNING This adjustment should only be performed by an instrument
service technician.
XXLO WCONCENTRATIONXXLO W
CO 15.80 PPM
SAMPLE 12:34 ALARMx
ANALOG OUTPUT CAL:
>VOLTAGE CHANNEL 1
VOLTAGE CHANNEL 2
VOLTAGE CHANNEL 3
VOLTAGE CHANNEL 4
VOLTAGE CHANNEL 5
VOLTAGE CHANNEL 6
CURRENT CHANNEL 1
ª
RANGE AVG DIAGS ALARM
CURRENT CHANNEL 2
CURRENT CHANNEL 3
CURRENT CHANNEL 4
CURRENT CHANNEL 5
CURRENT CHANNEL 6