EasyManuals Logo

Thermo NanoDrop One User Manual

Thermo NanoDrop One
312 pages
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Page #162 background imageLoading...
Page #162 background image
Measure Kinetics
156 NanoDrop One User Guide Thermo Scientific
To measure a sample using the Kinetics application
1. From the instrument Home screen, select the Kinetics tab and tap the Kinetics icon.
The Kinetics Setup screen is displayed. If one or more kinetic methods exist in the
currently selected Data Storage Location, they will be listed in the Select Method box. A
description of the selected method appears in the Method Details box.
2. Select a method:
select an existing method by tapping the method name in the Select Method box
create a new method by tapping Create Method, specifying the method settings
and choosing Save Method
edit an existing method by tapping the method name and choosing Edit Method
3. Specify any cuvette options such as heating or stirring by tapping > Settings (see
General settings for details).
Note: If your cuvette pathlength is not 10 mm, specify the correct pathlength in
General Settings.
4. Tap Run Method.
5. Measure a blank:
Fill clean, dry cuvette with enough blanking solution to cover instrument optical
path
Lift instrument arm and insert blanking cuvette into cuvette holder, making sure to
align light path of cuvette with light path of instrument
–Tap Blank

Table of Contents

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Thermo NanoDrop One and is the answer not in the manual?

Thermo NanoDrop One Specifications

General IconGeneral
Wavelength Range190-850 nm
Pathlength1 mm and 0.2 mm (Auto-ranging)
Detection Limit2 ng/µL (dsDNA)
Light SourceXenon flash lamp
Maximum Concentration15, 000 ng/µL (dsDNA)
Detector2048-element linear silicon CCD array
Measurement Time< 5 seconds
Weight2.0 kg (4.4 lbs)
Nucleic Acid QuantitationYes
Protein QuantitationYes
Display7-inch capacitive touch screen
Sample Volume1-2 μL
Absorbance Precision±0.002 A
Operating Voltage100 to 240 V AC
Dimensions (W x D x H)20 cm x 25 cm x 15 cm (8 in. x 10 in. x 6 in.)
SoftwareThermo Scientific™ software
ConnectivityUSB, Ethernet, Wi-Fi
Operating Temperature15 - 35°C
Dimensions20 cm x 25 cm x 15 cm (8 in. x 10 in. x 6 in.)
Absorbance Range0 - 550 A (10 mm equivalent)