Measure Kinetics
Thermo Scientific NanoDrop One User Guide 157
Related Topics
• Measure a Sample Using a Cuvette
If Heat Cuvette to 37 °C is selected in General Settings, a message tells you the
current temperature and waits for the heater to reach the target temperature before
starting the measurements:
– Wait for blank measurement to complete and then remove cuvette
Note: The heater target temperature is not adjustable.
6. Measure a sample:
– Fill clean, dry cuvette with enough sample solution to cover optical path
– Insert sample cuvette into cuvette holder, making sure to align light paths
–Tap Measure
If Heat Cuvette to 37 °C is selected in General Settings, a message tells you the
current temperature and waits for the heater to reach the target temperature before
starting the measurements:
Note: You may add reagents to the sample solution at any time during the
measurement
Use the Pause button at the bottom of the measurement screen to pause the
experiment (if you need to end the experiment early, tap Stop)
– Wait for all measurement stages to complete
– Remove cuvette and clean it according to manufacturer specifications
Results for each measurement in each interval are displayed in real time. When all stages
are completed, the spectra and reported values for the entire experiment are displayed.
7. When you are finished reviewing the data, tap End Experiment. Each saved experiment
contains one complete set of kinetic measurements based on the selected method.
To override the wait
and start the blank
measurement
immediately, tap
Blank Now.