EasyManua.ls Logo

Thermo NanoDrop One - Page 21

Thermo NanoDrop One
312 pages
Print Icon
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
Measure dsDNA, ssDNA or RNA
Thermo Scientific NanoDrop One User Guide 15
Best practices for nucleic acid measurements
Isolate and purify nucleic acid samples before measurement to remove impurities.
Depending on the sample, impurities could include DNA, RNA, free nucleotides,
proteins, some buffer components and dyes. See Preparing Samples for more information.
Ensure the sample absorbance is within the instrument’s absorbance detection limits.
Blank with the same buffer solution used to resuspend the analyte of interest. The
blanking solution should be a similar pH and ionic strength as the analyte solution.
•Run a blanking cycle to assess the absorbance contribution of your buffer solution. If the
buffer exhibits strong absorbance at or near the analysis wavelength (typically 260 nm),
you may need to choose a different buffer or application. See Choosing and Measuring a
Blank for more information.
For micro-volume measurements:
Ensure pedestal surfaces are properly cleaned and conditioned.
If possible, heat highly concentrated or large molecule samples, such as genomic or
lambda DNA, to 63 °C (145 °F) and gently (but thoroughly) vortex before taking a
measurement. Avoid introducing bubbles when mixing and pipetting.
Follow best practices for micro-volume measurements.
Use a 1-2 μL sample volume. See Recommended Sample Volumes for more
information.
For cuvette measurements (NanoDrop One
C
instruments only), use compatible cuvettes
and follow best practices for cuvette measurements.
Related Topics
Measure a Micro-Volume Sample
Measure a Sample Using a Cuvette
Best Practices for Micro-Volume Measurements
Best Practices for Cuvette Measurements
Prepare Samples and Blanks
Basic Instrument Operations
Note Extraction reagents such as guanidine, phenol, and EDTA contribute
absorbance between 230 nm and 280 nm and will affect measurement results if
present in samples (even residual amounts).

Table of Contents

Related product manuals