238 NanoDrop One User Guide Thermo Scientific
Dye/Chrom. Editor
Use the Dye/Chromophore Editor to add a custom dye to the list of available dyes in
Microarray Setup or Proteins & Labels Setup. You can also specify which dyes are available in
that list.
Protein Editor
Use the Protein Editor to add a custom protein to the list of available proteins in the
Protein A280 application.
Related Topics
Set Up the Instrument
Update Software
Measure a Sample Using a Cuvette
Set Up Ethernet Connection
Dye/Chromophore Editor
Protein Editor
Use cuvette Select cuvette sampling mode (available for NanoDrop One
C
instrument model only). When selected, these additional
options are available:
Pathlength: Enter cuvette pathlength (width) before taking
blank or sample measurements with cuvettes (see cuvette
manufacturer for cuvette specifications)
Stir Speed: If using automatic stirring, drop micro-stir bead
into sample cuvette and set Stir Speed (levels 1 through 9
correspond with range from 10 RPM to 850 RPM with
controlled ramping from zero)
Heat cuvette to 37 °C: Select this option if sample cuvettes
require heating. Cuvette heater increases from room
temperature to 37 °C at rate of 5 °C/minute.