ALADDIN - User manual Rev. 23 of 08/07/2016
41
The correction values (in mm) of the natural vitreous body
Phakic
0
APhakic
0.21
Pseudophakic Unknown material
0.11
Pseudophakic Silicone IOL
0.12
Pseudophakic PMMA IOL
0.11
Pseudophakic Acrylic IOL
0.1
Pseudophakic Memory IOL
0.11
For the vitreous body you can choose between:
Natural: the vitreous body has never been operated or treated such as to alter its composition.
Silicon Oil: the vitreous body has been filled, even only partly, with silicon oil.
The correction values (in mm) of the vitreous body filled by Silicon Oil
Phakic
-0.74
APhakic
-0.86
Pseudophakic Unknown material
-0.75
Pseudophakic Silicone IOL
-0.74
Pseudophakic PMMA IOL
-0.75
Pseudophakic Acrylic IOL
-0.76
Pseudophakic Memory IOL
-0.75
All this information is required because, on the basis of the artificial materials and their optical properties
present inside the eye, the instrument always corrects the measurements obtained to the most precise value
possible.
Once this information has been entered, you can access the acquisition environment.
For more details on the acquisition environment see the dedicated section.
The vitreous body nature is expressed, if different from natural, in the acquisition view as well as in the output
reports, as shown in the following figures. The lens nature is always reported.