EasyManua.ls Logo

Topcon ALADDIN - Performance Testing

Topcon ALADDIN
147 pages
Print Icon
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
ALADDIN - User manual Rev. 23 of 08/07/2016
137
The ACD measurement is the distance between corneal epithelium and the crystalline lens surface.
The light emitted from this instrument is NOT hazardous. It is classified ad Group 1 of ISO 15004-2.
15.4 Performance Testing
BRIEF SUMMARY OF PERFORMANCE TESTS AND RESULTS
A prospective, single site clinical study comparing the performance of the ALADDIN with IOL Master was
conducted in 63 eyes (1 eye for each enrolled subject). This study evaluated the agreement, repeatability and
precision in the subsequent endpoints:
Data Type
All Devices
Quantitative
Measurements
Axial Length
Anterior Chamber Depth
Radii of curvature of flattest and steepest meridian
Axis of the flat meridian
White to white distance
The agreement between instruments is summarized in Table 1:
Table 1.
Agreement between ALADDIN and IOL Master.
ALADDIN
IOL Master
Agreement parameters
Endpoint
Mean
SD
Mean
SD
Difference
Difference SD
lower LoA
upper LoA
acd
3.417830688
0.679989056
3.31952381
0.650987552
0.0983069
0.116363197
-0.1344195
0.33103327
al
23.70365079
1.703971867
23.68846561
1.710252709
0.0151852
0.027245176
-0.0393052
0.06967554
ax for 0.00<=cyl<=0.75
88.18518519
94.49566422
87.53947368
92.10832857
2.0765015
9.975699862
-17.874898
22.0279012
ax for 0.75<cyl<=1.50
77.48101266
115.8921525
76.05882353
114.2054185
0.9400949
3.261808332
-5.5835217
7.46371159
ax for
1.50<cyl
40.82758621
92.30826382
39.96428571
103.2499472
0.8218912
2.093718248
-3.3655453
5.00932767
k1
42.93549692
2.400645881
43.01523003
2.415238617
-0.0797331
0.154802285
-0.3893377
0.22987146
k2
43.90205198
2.589996595
43.99521136
2.600533074
-0.0931594
0.186267212
-0.4656938
0.27937504
wtw
12.28312169
0.525703699
12.32291005
0.557496798
-0.0397884
0.118840715
-0.2774698
0.19789307
Note to the Table 1:
- axis results are summarized by the subsequent classes of cylinder: 0.00cylinder0.75; 0.75<cylinder1.50; 1.50<cylinder.
- data in table concern the comparison producing the worst mean difference for al detected in the study.
The results of the study demonstrate that the axial length, corneal radius, anterior chamber depth and the
“white-to-white” measurements of the Aladdin are substantially equivalent to those of the predicate
device.
To evaluate repeatability and reproducibility, for each endpoint (AL, ACD, AX, K1, K2, WTW), analysis of
precision was performed for ALADDIN and IOLMaster separately using a factorial ANOVA models for repeated
measures within subject including the subsequent term: device identifier (A1, A2, A3 for ALADDIN and I1, I2,
I3 for IOLMaster), operator identifier (1, 2 and 3), subject identifier (1 to 63) and every two-way interactions;
a standard variance component matrix was used as covariance structure of R-side matrix. The related results

Table of Contents

Related product manuals