MC122 - MCP-4J Control Unit Section 3 - Basic features
27
MCP-4J Valve Driver Instructions
One of three wiring configurations can be used to trigger the valves. Chose the one that best fits the needs of your
system. Detailed instructions for each configuration are given in the next section.
WARNING! The triggering device used must be configured to output only 24VDC. If the control unit has a
spike, it will damage the input circuit for the scanner.
Wiring Configurations for Triggering the Valves
WARNING! Disconnect all power and input signal wiring before attempting to install or replace any part of
the glue system. Otherwise, personal injury or death may occur!
Wiring the Scanner Cable
1. Make sure the MCP-4J power switch is OFF and the unit is unplugged.
2. Cut the female connector off of the Scanner 1 cable.
3. Plug the male connector (on the Scanner 1 cable) into the Scanner 1 input port on the back of the MCP-
4J Control Unit.
4. Carefully strip the scanner cable jacket back and strip the individual wire insulations about 0.25 inch.
5. Wire the modified scanner cable according to the Wiring Diagram (next page).
Scanner cable (male connector) plugs into the Scanner 1 input port

Table of Contents

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the VALCO MELTON MCP-4J and is the answer not in the manual?

VALCO MELTON MCP-4J Specifications

General IconGeneral
BrandVALCO MELTON
ModelMCP-4J
CategoryController
LanguageEnglish