PaX-i (PCH-2500) User Manual
44
4. Software Overview
G. Patient Gender
This displays the current patient’s gender as entered in EasyDent’s patient
information. If necessary, gender can be manually selected.
Age Group / Gender VATECH’s Standard
Child
≤ 12
Adult
Man
≥ 13
Woman
H. X-ray intensity
This tool selects X-ray intensity.
NOTE
Depending on the circumference of the patient's head,
X-ray intensity may be classied as Hard, Normal, or soft:
Soft ≤ Normal ≤ Hard
Age
Group
Average head
circumference
Range
(cm)
Classication
of head
circumference
Child 53±3 cm
> 53±3 Hard
53±3 Normal
53±3 < Soft
Adult 56±3 cm
> 56±3 Hard
56±3 Normal
56±3 < Soft
I. Phantom capture
This function is used when the Phantom Jig
is being used to acquire images.
Pi_u272_en_20180514.indd 44 2018-05-14 오전 11:03:16

Table of Contents

Other manuals for Vatech PAX-I

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Vatech PAX-I and is the answer not in the manual?

Vatech PAX-I Specifications

General IconGeneral
Focal Spot Size0.5 mm
Image ReceptorDigital sensor
CompatibilityDICOM 3.0
Image DetectorCMOS
Power Supply220-240 VAC, 50/60 Hz
Sensor TypeCMOS
X-ray GeneratorHigh Frequency
Tube Voltage60-90 kVp
Tube Current4-12 mA
Power Requirements100-240V AC, 50/60Hz
UsageDental diagnostics