VEMCO – VR100 Manual 28 May 2018 55
STEP 7
Wait the selected amount of time while noise samples are taken.
For the example above, that would be ten minutes.
STEP 8
Press “Menu” to return to the Noise menu list. Select “Stop & Read” (#2) to
stop recording the noise values and retrieve the information.
NOTE: HR2 only - an additional screen prompts for RMS, PPM, or HR noise.
STEP 9
Wait while the VR100 retrieves the noise information from the receiver.
STEP 10 (VR2AR screen shown)
The minimum noise level (in mV), the average noise level (in mV), and the
maximum noise level (in mV) are all listed in the order.
NOTE: The HR2 reports these in dBs.
Press resend to repeat the retrieval process, if desired.
STEP 11
Use the MENU button to return to the main transponding screen.
VR2AR:000001 is now
Sampling noise.
Ok
1 Setup & Start...
2 Stop & Read
Select
VR2AR:000001
Reading noise...
T-15 76dB
▊▊▊
VR2AR:000001 170 m
Min: 190 mV
Avg: 210 mV
Max: 230 mV Resend