Maintenance and Troubleshooting Yamatake Corporation
7-24 ST3000 Smart Transmitter Series 900 Electronic Differential Pressure/Pressure Transmitter
7-4-2-4 Saving Calibrated Values and Releasing Constant Current
Source Mode
Save the calibrated value to non-volatile memory of this transmitter and release the
constant current mode:
Step Description SFC screen
1
Press the key.
2
Press the keys.
Forceful data saving is com-
pleted.
3
Press the key.
The current output is 100%
4
Press the key.
# disappears, indicating that the
constant current mode has been
released.
SHIFT
SHIFT- #
ENTER
( Yes )
NON-VOL
DSTJ TAG NO.
WORKING...
DSTJ FIT-1234
DATA NONVOLATILE
DSTJ FIT-1234
READY...
OUT-
J
PUT
INPUT
OUTPUT FIT-1234
WORKING...
OUTPUT FIT-1234
100.00 % #
CLR
( No )
OUTPUT FIT-1234
WORKING...
DSTJ FIT-1234
READY...