Saving and Loading Data
• Extension and Data Size
Data Type Extension Data Size (Bytes)
Binary .wvf Approx. (8K + 4 × the number of waveform sampling
data points × Tw) where Tw is the number of displayed
waveforms
.hdr Approx. 1 K (condition: when there is one displayed
waveform)
ASCII .csv Approx. 150 K (condition: when there is one displayed
waveform) with the data update rate set to 50 ms
Float .fld 4 × the number of waveform sampling data points
• The number of waveform sampling data points of U1 to I4, Speed, Torque, Math1,
Math2 in waveform computation mode
Data Update Rate Number of Waveform Sampling Data Points
50 ms 10,000 points
100 ms 20,000 points
250 ms 50,000 points
500 ms 100,000 points
1 s 200,000 points
2 s 400,000 points
5 s 1,000,000 points
10 s 2,000,000 points
20 s 4,000,000 points
• The number of waveform sampling data points of U1 to I4, Speed, Torque, Math1,
Math2 in FFT mode
FFT Points Number of Waveform Sampling Data Points
200 k 200,000 points
20 k 20,000 points
* The number of waveform sampling data points do not depend on the data update rate.
• The number of waveform sampling data points of FF1 and FF2 in FFT mode
FFT Points Number of FFT Waveform Data Points
200 k 100,000 points + 1 (DC)
20 k 10,000 points + 1 (DC)
• Selecting the Waveform
• When the data type is set to Binary or ASCII, the waveform that is displayed on
the screen is saved.
• When the data type is Float, the waveform of the input signal selected from the
following is saved.
• Only the input signals of the installed elements are applicable for the
selection.
• For models with the motor evaluation function (motor version), the input
signals of Speed and Torque are also selectable items.
9.3 Saving Setup Parameters, Waveform Display Data, Numeric Data, Waveform Sampling Data

Table of Contents

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the YOKOGAWA wt3000 and is the answer not in the manual?

YOKOGAWA wt3000 Specifications

General IconGeneral
Frequency RangeDC, 0.1 Hz to 1 MHz
Number of Input ElementsUp to 4 elements
Harmonic MeasurementUp to 500th order
InterfacesUSB, Ethernet, RS-232
Sampling Rate2 MS/s
Voltage Range1.5 V to 1000 V
Display8.4-inch LCD