Page #265 background image
9-18 IM 760301-01E
Executing the Save Operation
The save operation may take several seconds to several hours depending on
the number of waveforms to be saved, the data type, the number of waveform
sampling data points determined by the data update rate, and the data transfer
rate to the save destination medium. The save operation takes a long time when
there are many waveforms to be saved, the data update rate is slow, and the
number of waveform sampling data points is large.
The measurement stops while the data is being saved. The measurement
resumes when the save operation completes or is aborted.
To execute the save operation when the data update rate is 20 s and the display
is held, execute a single measurement once, and carry out the save operation
after the data update due to the single measurement is complete.
If you carry out single measurement while the display is held, the waveform
sampling data is overwritten. If you execute the save operation while the data
update by the single measurement is in progress and the waveform sampling
data is being overwritten, an error occurs, and the file is not created.
If the speed or torque signal type of the motor evaluation function is set to pulse,
“ERROR” is saved for the waveform sampling data of that signal.
Selecting the Storage Medium and Directory
Media on which saving and reading is possible are displayed in the File List dialog
box.
Currently selected medium and directory
Directory
Save destination medium
Display Example of Storage Media
[PC_Card]: PC card
[USB2]: USB memory (option) with ID number 2
[NetWork]: Network drive (option. see
Expansion Function User’s Manual IM760301-
51E
)
9.3 Saving Setup Parameters, Waveform Display Data, Numeric Data, Waveform Sampling Data

Table of Contents

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the YOKOGAWA wt3000 and is the answer not in the manual?

YOKOGAWA wt3000 Specifications

General IconGeneral
Frequency RangeDC, 0.1 Hz to 1 MHz
Number of Input ElementsUp to 4 elements
Harmonic MeasurementUp to 500th order
InterfacesUSB, Ethernet, RS-232
Sampling Rate2 MS/s
Voltage Range1.5 V to 1000 V
Display8.4-inch LCD