OPERATION
Carl Zeiss Coded Axio Imager operation and function controls Axio Imager
80 430000-7344-001 M70-2-0020 e 06/2009
Intensity button (
4-1/34)
− Reduces the light intensity of halogen illuminator HAL.
Intensity button (
4-1/35)
− Increases the light intensity of halogen illuminator HAL.
Toggle switch for transmitted/reflected-light halogen illuminator (
4-1/36)
− Alternately switches the halogen illuminator for transmitted light or reflected light on/off.
− Toggle switch up: Reflected-light halogen illuminator ON (transmitted-light OFF)
− Toggle switch down: Transmitted-light halogen illuminator ON (reflected-light OFF)
Button LM-Set (
4-1/37)
− Saves the values of the current light intensity and the optical path nosepiece position (brief beep).
− Adaptation of the luminosity of the ring-like arranged LEDs at the light intensity control:
− Switch on the illuminator and adjust light intensity at the control so that all LEDs light up.
− Maintain LM-Set key pressed down. You hear a long beep tone.
− Adjust the desired light intensity using the light intensity control (4-1/14).
− Release the LM-Set key to save the light intensity adjusted.
Lever for reflected-light diffusing glass (
4-1/38)
− Lever in top position: Diffusing glass in light path
− Lever in bottom position: Diffusing glass removed from light path