2.8 Technical data
Dimensions (width x depth x height)
Axio Imager stand, manual with HBO 100 ...................................... approx. 300 mm x 721 mm x 505 mm
Axio Imager stand, motorized with HBO 100 and TFT display.......... approx. 390 mm x 721 mm x 505 mm
Weight
Axio Imager, manual/motorized (dependent on equipment)........................................ approx. 18 to 40 kg
Ambient conditions
Transport (in packaging):
Permissible ambient temperature ......................................................................................... -40 to +70 °C
Storage:
Permissible ambient temperature ......................................................................................... +10 to +40°C
Permissible relative humidity (no condensation)............................................................ max. 75% at 35 °C
Operation:
Permissible ambient temperature ........................................................................................ +10 to +40 °C
Permissible relative humidity ........................................................................................max. 75% at 35 °C
Atmospheric pressure ............................................................................................... 800 hPa to 1060 hPa
Altitude ................................................................................................................................ max. 2000 m
Pollution degree ......................................................................................................................................2
Operating data for coded Axio Imager, equipped with an integrated power supply or
motorized Axio Imager using the VP232-2 external power supply
Operating environment.......................................................................................................... Closed room
Protection Class........................................................................................................................................ I
Protection Type.................................................................................................................................. IP 20
Electrical safety.............................................................in compliance with DIN EN 61010-1 (IEC 61010-1)
including CSA and UL directives
Overvoltage category............................................................................................................................... II
Radio interference suppression .........................................................in accordance with EN 55011 Class B
Noise immunity................................................................................in accordance with DIN EN 61326 /A1
Line voltage for integrated power supply .........................................100 to 127 and 200 to 240 V ±10 %
Change of line voltage setting is not required!
Line voltage for external power supply VP232-2.........................................................100 to 240 V ±10 %
Line frequency.............................................................................................................................50/60 Hz
Power consumption of coded Axio Imager.............................................................................max. 260 VA
Power consumption of Axio Imager, motorized......................................................................max. 190 VA
LED illuminator ........................................................................................ 400 to 700 nm, peak at 460 nm
Attachment lamp VIS-LED........................................................................ 400 to 700 nm, peak at 460 nm

Table of Contents

Related product manuals