Starting up
/
Switching off the leak
detector
Starting up of the leak
detector
(continued)
When the
pumps
have reached their nominal speed, the unit auto-
calibrates: different
screens
will show.
1
When calibration is completed,
the unit is ready to start a cycle.
...
STDBY
The
digital voice gives to the operator the message:
“detector ready for
cycle”
.
Flashing
When the filament is OK, the unit
is ready to start a cycle.
A « i » is flashing at the right end
of the
1st
line of the LCD.
READY FOR CYCLE
...
...
i
...
It
informs
the operator that the
unit is in manual calibration:
2
it will remain in manual
calibration until operator
performs
an
auto-calibration.
STDBY
Information message
C
580
Internal calibration on request
C
300
Note: It is possible to start a
test
cycle even if the
detector
is not
calibrated.
Switching
of
f
the leak detector
The
leak
detector
can be switched off at any time by
depressing
the
main power switch to 0
(OFF)
but it is
recommended (especially
when
the leak
detector
may not be
used
for an extended period of time) to
stop it following the procedure described below
.
2