EasyManua.ls Logo

Adixen ASM GRAPH D+ - D 100 - Table of preventive maintenance intervals

Default Icon
336 pages
Print Icon
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
D
100
T
able
of preventive
maintenance
intervals
ASM
102
S
ASM
122 D
ASM
142
ASM
142
S
ASM
142 D
ASM
Graph D+
ASM
182
T
ASM
192
T
ASM
192 T2
ASM
182 TD+
ASM
192 TD+
ASM
192 T2D+
ASM
1002
ASI
22
1000 h
(1)
Clean
filters
(inlet
filters,
air inlet
filter)
-
2000 h
(1)
or
3 months
(2)
Change the rotary vane pump
oil.
Change the oil mist eliminator.
E
750
4000 h
(1)
or
6 months
(2)
Clean the
vacuum
lines, the valves
and the gauges with alcohol
-
Dust
the electronic boards and the fans.
Partial
maintenance
of the analyzer
cell:
Replace
analyzer cell
filaments
and collector. Clean the
analyzer cell with alcohol (this
cleaning may be
necessary
in
case
of general internal contamination
creating insulating deposits).
E
400
8000
h
(1)
ou 1 year
(2)
Sniffer probe filter
replacement
if
used.
G
200
Pirani gauge adjustment.
CS
APT
1004 gauge adjustment.
CS
Electronic
overhaul
-
Replace
the seal in the rotary vane
pump.
CS
AMD1 pump:
replacement
of
membranes
and check valves.
Change the inlet filter of the test
chamber.
E
710
GB 02469 -
Edition
03 -
June
08
Frequency*
Maintenance
operations to be performed
CS
:
Please
contact
Customer
Service
*Service
intervals: The service intervals
given are for applications and work
rates
which conform to the normal operating
conditions. If the
machine
is operating under more difficult conditions they can be shortened.
(1) running time
(2) running time or storage
(3) storage
1/2

Table of Contents

Related product manuals