740
Greasing molecular and
turbomolecular pumps
GB 01240
-
Edition 04
- Juin
08
«Running in»
the pump after
relubrication
On the
P0326/P0365
converter
boards:
After relubricating the pump, the grease
must
be distributed in the
bearings
(“Running
in”).
This
is performed by limiting the speed of the pump for a
short
period.
For this, on the
P0326/P0365
converter
boards,
set
the running in
switches;
switch 1 to ON
(detector
off).
Start up the
detector
and let it operate for approximately 10
minutes
in this position.
After 10
minutes,
switch off the
detector
and
reset
switch 1 to
OF
F
.
Start the
detector
up again and check that the pump
reaches
its
synchronization
speed correctly. If this time is too long, repeat the
running in procedure a
second
time.
P0326 board P0365 board
ON ON
1
2
Normal
position
1
2
Running
in
position
3/5