Installation
of
a new internal
calibrated leak
After recalibration, your «recalibrated» leak is
returned
to you with:
-
its new value,
-
a calibration certificate.
To
install the calibrated internal leak (new or recalibrated):
Switch off the
detector
and unplug it from the main power
.
Install
the leak:
-
connect
the flange,
-
connect
the heat sensor.
Entering
the new internal calibrated leak values.
Calibrated leak
values
programming
C
305
Close the leak detector.
Connect
the
detector
to the main power
.
Switch it on: a calibration is performed automatically at the end of
the
start-up sequence
if the autocal is on.
It is
recommended
to repeat a calibration after 1 hour of operation
when the
temperature
has stabilized inside the unit.
Basic
internal calibration of the leak detector
C
301
Intensive use
of the
detector
In the case of
intensive use
of the detector, it is
recommended
to have
a spare internal calibrated leak.
If this is not possible, the
detector
can still be
used
and calibrated
using an external calibrated leak
.
External
calibration of the leak detector
C
303
2
/