Adixen ASM GRAPH D+ - Installation of a New Internal Calibrated Leak; Intensive Use of the Detector

Default Icon
336 pages
Print Icon
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
E
560
Replacement
of the internal calibrated
leak
GB 01253 -
Edition
04 -
June
08
Installation
of
a new internal
calibrated leak
After recalibration, your «recalibrated» leak is
returned
to you with:
-
its new value,
-
a calibration certificate.
To
install the calibrated internal leak (new or recalibrated):
Switch off the
detector
and unplug it from the main power
.
Install
the leak:
-
connect
the flange,
-
connect
the heat sensor.
Entering
the new internal calibrated leak values.
Calibrated leak
values
programming
C
305
Close the leak detector.
Connect
the
detector
to the main power
.
Switch it on: a calibration is performed automatically at the end of
the
start-up sequence
if the autocal is on.
It is
recommended
to repeat a calibration after 1 hour of operation
when the
temperature
has stabilized inside the unit.
Basic
internal calibration of the leak detector
C
301
Intensive use
of the
detector
In the case of
intensive use
of the detector, it is
recommended
to have
a spare internal calibrated leak.
If this is not possible, the
detector
can still be
used
and calibrated
using an external calibrated leak
.
External
calibration of the leak detector
C
303
2
/
2