Remote
control
Remote
control
connecting
In order to
use the remote control
with
control panel
with graphic
interface,
it is
necessary
to
connect the remote control before
starting
up
the
detector.
Location
and connecting
B
210
Remote
control choice
3 different
units
could be
selected
in the leak
detector
but only one
unit is available on the
remote
control.
So the operator should choice its
remote
control in accordance of the
operation unit chosen.
Unit
C
570
Accessories
A
700
When the operator connects the remote control
on
the
leak detector,
the
leak
detector
unit is
automatically reprogrammed
with
the
unit
of
the remote control. The remote control
unit is
memorized
by the
detector when the operator disconnects the remote
control.
Use
and display
CYCLE
Analog and digital displays
To
start/stop a cycle
The remote
control:
•
allows to display leak
measured
value,
•
allows to start/stop cycle, zero function and internal calibration,
•
allows to display
status
of the air inlet vent, sniffing
test
mode and
external calibration,
•
doesn’t allow to adjust leak
detector
parameters.
On
remote
control and control panel, the displayed
values
on the
analog and digital displays are exactly the same.
Operation of the leak detector
C
211
In order to start/stop a cycle, operator can
use
either the CYCLE
control key on the control panel or
remote
control.
Display
The LED
indicator is ON/OFF when the control key is
activated/
deactivated.
As soon as the
detector
is in
test,
the
TEST
indicator is
ON.
2/3